Orchestra® Module DPS Visio


Orchestra® Module DPS Visio
The Module DPS VISIO can be integrated into all Orchestra® Workstation. One syringe drive, connectable to Base A, Base Primea and Base Intensive.
Features
- Rate range: from 0.1 to 1200 ml/h
- 5, 10, 20, 30/35, 50/60 ml. automatic syringe volume recognition
- Flow rate accuracy ± 1% on drive mechanism/ ± 2% on syringes
- Choice of infusion modes
- Volume per hour mode
- Mass infusion units mode
- Drug Library up to 50 drugs with corresponding concentrations
- Bolus injection “hands free” bolus, programmable in amount and flow rate, calculated in dose units or not. Manual bolus, programmable up to 1200 ml/h
- Occlusion and pressure management easy setting of occlusion alarm trigger level, up to 1100 mmHg
- Line disconnection alarm drop in pressure in the infusion line
- Dynamic pressure system (DPS) increase and decrease of the pressure in line monitoring
- Anti-bolus system bolus volume reduction at occlusion release
- Volume limit and total volume
- Battery life: 7 hours minimum, at 5 ml/h
- Dimensions/Weight W=130, H=315, D=105 mm/2.3 kg
Technical Data
Please view the technical data in the PDF below.
PDF-Download (1.8 MB)
Service
Fresenius Vial Competence Center
Fresenius Vial
Le Grand Chemin
38000 Brezins
FRANCE
| E-mail: | contact.vial@fresenius-kabi.com |
International Hotline & Training Service
| E-mail: | maintenance.vial@fresenius-kabi.com |
Quality & Regulatory Affairs
Quality Service
Marie Pierre Pruneau
| Phone: | +33 476 67 1040 |
| Fax: | +33 476 67 1093 |
| E-mail: | qa-ra.vial@fresenius-kabi.com |
Safety Officer
Michel Monier
| Phone: | +33 476 67 1041 |
| Fax: | +33 476 67 1093 |
| E-mail: | qa-ra.vial@fresenius-kabi.com |
Regulatory Files
Alexia Garin
| Phone: | +33 476 67 1159 |
| Fax: | +33 476 67 1093 |
| E-mail: | qa-ra.vial@fresenius-kabi.com |